General Specifications for measure
These specifications assume:
A 1-year calibration cycle
An operating temperature of 18 to 28
Relative humidity of 35% to 70% (non_condensing)

Rate:

Measurement function Rate
DCVDCIOHMTC 2 Readings per Second about
RTD 1 Readings per Second about
FREQ  0.5 Readings per Second about
CONT. 4 Readings per Second about
PRESSURE 1.5 Readings per Second about
 
 
DCV
Normal Mode Rejection Ratio (NMRR) ≥60dB (at 50Hz or 60Hz)
Common Mode Rejection Ratio (CMRR) ≥140dB (at 50Hz or 60Hz)
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C for 5 to 18 and 28 to 40
The range of the internal temperature compensation sensor is from -10 to 50, compensation error0.5
Maximum voltage between VΩHz terminal and COM terminal: 60 Vpk
Maximum Input current: 60mA. 
Protected with a 63mA, 250V fast blow fuse
General Specifications for Source
These specifications assume:
A 1-year calibration cycle
An operating temperature of 18 to 28 (64.4~82.4)
Relative humidity of 35% to 70% (non_condensing)


Measurement and source pressure
 
PRESSURE
Rang and  
Accuracy
resolution For more detail, 
refer the pressure 
module about APM.
Determined 
by 
pressure 
module
5 digits
Other feature:
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C for 5 to 18 and 28 to 40.
The range of the internal temperature compensation sensor is from -10 to 50
Maximum voltage between any output terminal and earth: 30Vpk
Maximum output current: Approximately 25mA.

Function Reference Range Resolution Accuracy Remark
DCV 50mV -5.000mV
55.000mV
1μV 0.02+0.02 Input Resistance:
 100MΩ
500mV -50.00mV
550.00mV
10μV 0.02+0.01
5V -0.5000V
5.5000V
0.1mV 0.02+0.01 Input Resistance:
 1MΩ
50V -5.000V
55.000V
1mV 0.03+0.01
DCmA 50mA -5.000mA
55.000mA
1μA 0.02+0.01 Shunt Resistance:10Ω
OHM 500Ω
Test
Current:
Approxim
tely 1mA
0.00Ω550.00Ω 0.01Ω 0.05+0.02 Open Circuit Voltage
about 2.5V; 
Does not include lead 
resistance;
5KΩ
Test
Current:
Approxim
tely 0.1mA
0.0000 KΩ
5.5000KΩ
0.1Ω 0.05+0.02
FREQ 500Hz 3Hz500.00Hz 0.01Hz ±2digit Input Impedance
100 kΩ at least;
Sensitivity:3Vp-p 
minimum;
Duty Cycle: 50%.
5KHz 3Hz
5.0000KHz
0.1Hz
50KHz 3Hz50.000KHz 1Hz
TC R 0°C1767°C 1°C 0500℃: 1.8℃
5001767℃  : 1.5℃
By using ITS-90 
temperature scale;
The accuracy does
not include the error 
of internal 
temperature 
compensation 
caused by a sensor;
 
S 0°C1767°C
K -100.0°
1372.0°C
0.1°C -100.00.0℃: 1.2℃
0.0~1372.0℃: 0.8℃
E -50.0°C
1000.0°C
  -50.0℃0.0℃: 0.9℃
0.0~1000.0℃: 1.5℃
J -60.0°C
1200.0°C
-60.00.0℃: 1.0℃
0.0~1200.0℃: 0.7℃
 
T -100.0°C
400.0°C
-100.00.0℃: 1.0℃
0.0~400.0℃: 0.7℃
N -200.0°
1300.0°C
-200.00.0℃: 1.5℃
0.0~1300.0℃: 0.9℃
B 600°C
1820°C
1°C 600800℃: 2.2℃
8001000℃: 1.8℃
10001820℃: 1.4℃
RTD Pt100
385
-200.0°C
800.0°C
0.1°C -200.00.0℃ : 0.5℃
0.0400.0℃ : 0.7℃
400.0800.0℃ : 0.8℃
By using Pt100-385
Does not include lead 
resistance.
Pt1000
385
-200.0°C
630.0°C
-200.0100.0℃: 0.3℃
100.0300.0℃ : 0.5℃
300.0630.0℃ : 0.7℃
Pt200
385
-200.0°C
630.0°C
-200.0100.0℃: 0.8℃
100.0300.0℃ : 0.9℃
300.0630.0℃ : 1.0℃
Pt500
385
-200.0°C
630.0°C
-200.0100.0℃: 0.4℃
100.0300.0℃ : 0.5℃
300.0630.0℃ : 0.7℃
Cu10 -100.0°C
260.0°C
1.8
Cu50 -50.0°C
150.0°C
0.7℃
SWITCH   CLOSE / 
OPEN
    Approximately 1mA 
Test Current
Short circuit display : 
CLOSE
Open circuit display:
OPEN
Threshold value about 
200~300Ω
CONT. 500Ω ≤50Ω sound     Approximately 
1mA Test Current
Function Reference Range Resolution Accuracy Remark
DC 
voltage
 
100mV -10.000mV
110.000mV
1μV 0.02+0.01 Maximum output 
current: 0. 5mA
1V -0.10000V
1.10000V
10μV 0.02+0.01 Maximum output 
current: 2mA
10V -1.0000V
11.0000V
0.1mV 0.02+0.01 Maximum output 
current: 5mA
DC 
current
20mA 0.000mA
22.000mA
1μA 0.02+0.02 External supply for 
simulate mA: 5V–28V
Maximum load 1KΩ at 20mA
Resistance
 
400Ω 0.00Ω
400.00Ω
0.01Ω 0.02+0.02 Excitation current: ± 0.5–3 mA;
if ± 0.1–0.5, add 0.1Ω;
Accuracy does not include
 lead resistance;
4KΩ 0.0000 KΩ
4.0000 KΩ
0.1Ω 0.05+0.025 Excitation current: 
±0.05 0.3mA;
Does not include
 lead resistance;
40KΩ 0.000 KΩ
40.000 KΩ
0.1+0.1 Excitation current: 
±0.01mA;
Does not include 
lead resistance;
TC
 
R 0°C1767°C 1°C 0100℃ : 1.5℃
1001767℃:  1.2℃
By using ITS-90 
temperature scale;
The accuracy does 
not include the error 
of internal 
temperature compensation 
caused by a sensor;
S 0°C1767°C 0100℃ : 1.5℃
1001767℃:  1.2℃
K -200.0°C
1372.0°C
0.1°C -200.0-100.0 : 0.6℃
-100.0400.0℃:0.5℃
400.01200.0℃: 0.7℃
1200.01372.0 :0.9℃
E -200.0°C
1000.0°C
-200.0-100.0 : 0.6℃
-100.0600.0℃:0.5℃
600.01000.0℃: 0.4℃
J -200.0°C
1200°C
-200.0-100.0 : 0.6℃
-100.0800.0℃:0.5℃
800.01200.0℃: 0.7℃
T -250.0°C
400.0°C
-250.0400.0℃: 0.6℃
N -200.0°C
1300.0°C
-200.0-100.0℃:1.0℃
-100.0900.0℃:0.7℃
900.01300.0℃:0.8℃
B 600°C
1820°C
1°C 600800℃ :  1.5℃
8001820℃:  1.1℃
RTD Pt100-385 -200.0°C
800.0°C
0.1°C -200.00.0℃ :  0.3℃
0.0400.0℃ :  0.5℃
400.0800.0℃: 0.8℃
By using Pt100-385
Excitation current: 
±0.5±3mA for Pt100, 
Cu10, Cu50;
Excitation current: 
±0.05mA±0.3mA 
for PT200, PT500, PT1000;
Does not include 
lead resistance.
Pt200-385 -200.0°C
630.0°C
-200.0100.0℃: 0.8℃
100.0300.0℃ : 0.9℃
300.0630.0℃ : 1.0℃
Pt500-385 -200.0°C
630.0°C
-200.0100.0℃: 0.4℃
100.0300.0℃ : 0.5℃
300.0630.0℃ : 0.7℃
Pt1000-385 -200.0°C
630.0°C
-200.0100.0℃: 0.2℃
100.0300.0℃:  0.5℃
300.0630.0℃:  0.7℃
Cu10 -100.0°C
260.0°C
-100.0260.0℃ :  1.8℃
Cu50 -50.0°C
150.0°C
-50.0150.0℃ :  0.6℃
FREQ 100Hz 1.00Hz
110.00Hz
0.01Hz ±2 count Output voltage: 
+1~+11 Vp-p 
(zero base waveform);
Amplitude accuracy:
 ±(5% +0.5V);
Maximum load: >100 KΩ;
Duty Cycle: 50%.
1KHz 0.100KHz
1.100KHz
1Hz
10KHz 1.0KHz
11.0KHz
0.1KHz
100KHz 1KHz
110KHz
2KHz ±5 count
PULSE 100Hz  1
100000cycles
1cyc ±2 count
1KHz
10KHz
SWITCH 100Hz 1.00Hz
110.00Hz
0.01Hz ±2 count FET switch
Maximum open/close
 voltage: +28 V
Maximum open/close current:: 
50mA
1KHz 0.100KHz
1.100KHz
1Hz
10KHz 1.0KHz
11.0KHz
0.1KHz
100KHz 10KHz
110KHz
2KHz ±5 count
LOOP 24V     ±10% Maximum current: 22 mA
Short circuit protected